Mindray A8 is the newest, state-of-the-art anesthesia machine that introduces many advanced features:
- Volume Exchanger (VE) breathing circuit that is more compact and precise than the traditional bellow.
- 18.5 inch touchscreen with 340-degree rotating arm.
- 8.4 inch auxiliary monitoring screen.
- ICU-level ventilation mode: APRV and AMV.
- Full range of monitoring option: CO2, Anesthesia gas, BIS, NMT, Transpulmonary pressure monitoring.
Detailed specifications:
- Dimensions (mm): 1490 x 910 x 705.
- Weight: ~160kg.
- 18.5 inch color touch screen, 1920 * 1080 resolution, 360 degree rotation, 60 degrees front and back tilt.
- 8.4-inch color secondary display. Displays gas exchange indicator, gas supply pressure and AGSS unit status.
- Dimensions of working space (mm): 850 x 590 x 325. Load capacity 30kg.
- Wheel diameter: 150mm, with central locking pedal.
- There are 3 drawers with lock, size 135mm high, 440mm wide, 385mm deep.
- Standard rail GCX, load bearing 27kg at a distance of 0.41m, 130mm x 130mm x 485 mm.
- The arm of the breathing bag has a size of 1130 * 312mm, which can be rotated 120 degrees.
- LED worklight: Off, Low, High.
- Power supply:
- 220-240 Vac, 50/60 Hz, 10A max.
- 100-240 Vac, 50/60 Hz, 10A max
- 100-120 Vac, 50/60 Hz, 10A max
- AC power cord length: 5m.
- Backup battery: Li-Ion, 12 VDC, 16Ah (or 32Ah 2 batteries). Use at least 90 minutes / 1 battery. 180 minutes / 2 batteries.
- Charging time: <12h.
- Output of auxiliary power connector: 3A/socket. Or 5A for 4 outlets.
- Supported connection ports: RS-232C, 2 RJ-45 (LAN) ports, 4 USB ports.
- Allowable operating temperature: 10 ~ 40°C.
- Allowable operating humidity: 15% ~ 95% (noncondensing).
- Permissible operating pressure: 70 ~ 106.7 kPa.
- Waterproof and dustproof standard: IPX1.
Anesthesia machine parameters:
- Support 2 vaporizer ports, Selectatec / Plug-in standard.
- Anesthesia: Halothane, Isoflurane, Sevoflurane, Desflurane.
- Measurement range of anesthetic gas consumption: 0-3000ml, accuracy ±2 mL or ±15%.
- Anesthesia vaporizer rate:
- Desflurane: 0 ~ 900 ml/h
- Sevoflurane: 0 ~ 450 ml/h
- Enflurane, Isoflurane and Halothane: 0 ~ 250ml/h.
- Accuracy: ±2ml/h or ±15% of measured value, whichever is greater.
- Flowpause function adjustable 1-2 minutes.
- Volume of breathing block: 1800ml (ventilator) or 1950ml (manual).
- CO2 absorber capacity: 1500ml or 1 Pre-pak.
- Supports 2 types of absorbers: Pre-pak or Loose-Fill.
- There is a removable 6ml water trap for the CO2 absorber.
- Inspiratory pressure measurement limit: -2 ~ 100 cmH2O. Accuracy: ± 2% FS.
- 2 Flow sensor type differential pressure at the inspiratory and expiratory ports.
- Galvanic cell Oxygen sensor, FiO2 18%-100%. ± 2.5%. Response ≤ 20s.
- Diameter of connectors: 22mmOD/15mmID. Anesthesia outlet port: 30mm OD
- Switch to switch 2 modes of mechanical ventilation – bag breathing.
- Valve APL 5~75 cmH2O. Adjustment rotation 330 degrees ± 10. Feedback ≥ 30 cmH2O.
- The system keeps the breathing system temperature stable at 35°C (when the machine is set at 20°C).
- All parts of the exhalation end are autoclavable (except the sensors).
- Compliance of the whole respiratory system: ≤ 2 mL/cmH2O ~ ≤ 4 mL/cmH2O.
- Breathing system impedance: ≤ 6 cmH2O (both Automatic and Manual mode).
- System leakage: ≤ 150 mL @ 3 kPa.
- Safety pressure limit on breathing line: 110 ± 10 cmH2O @ 10 – 110 L/min.
Ventilation settings:
- Patient Size: Adult, Child, Neonate.
- Tidal volume:10~1500 mL (volume mode) or 5~1500 mL (pressure mode). TV/IBW display is available.
- Pinsp: 5~90 cmH2O.
- Plimit: 10~100cmH2O.
- ∆Psupp 3~60 cmH2O.
- Respiration Rate: 4 ~ 100 bpm.
- I:E: 4:1 ~ 1:8.
- Tpause: OFF, 5% – 60%.
- Tinsp: 0.2 – 10.0 s.
- Trigger window: 5% – 90%.
- Flow trigger: 0.2 ~ 15 L/min.
- Pressure trigger: -20 ~ -1 cmH2O.
- Exp%: 5% – 80%.
- Min Rate: 2 – 60 bpm.
- Tslope: 0.0 – 2.0 s.
- Apnea I:E: 4:1~1:8.
- ∆Papnea: 3 – 60 cmH2O.
- Phigh: 3 to 90 cmH2O.
- Plow: 3 to 50 cmH2O.
- Thigh: 0.2 to 10.0 s.
- Tlow: 0.2 to 10.0 s.
- MV%: 25% to 350%.
- PEEP: OFF, 0~50 cmH2O. Electronic Control.
- Pressure limit of central air source: 280 to 600 kPa. Standard DIN/NIST gas head.
- Inlet gas pressure limit: Air/Oxygen: 6.9 ~ 20 Mpa; N2O:4.2 ~ 6 Mpa.
- Maximum flow rate: 180L/min + Fresh gas flow.
- Oxygen supply failure alarm: ≤ 220 kPa.
Monitoring parameters:
- Minute Volume: 0~100 L/min.
- Tidal volume:0 ~ 3000 mL/min.
- FiO2: 18% – 100%.
- Ppeak/ Pmean/ Pplateau: -20 ~ 120 cmH2O.
- PEEP: 0 ~ 70 cmH2O.
- I:E: 4:1 ~ 1:10
- Resitance (R): 0 ~ 600 cmH2O/(L/s).
- Compliance (C): 0 ~ 300 ml/cmH2O.
- Rate: 0 ~ 150 bpm.
- Trend: 48 hours.
- Screenshots: up to 50 photos.
- Event log: 10000 events.