- Description
Description
Equipped with smarter ventilator technology and enhanced monitoring capabilities, the WATO EX-65 supports precise anesthesia and comprehensive ventilation modes, to fully meet the needs of all your patients.
- New PC-VG and CPAP/PS ventilation mode
- Electronic display Flowmeter.
- Optimizer function to indicate the optimal fresh gas flow for anesthesia.
- AA measurement function to calculate the end-case usage of anesthesia gas.
- AG and BIS module.
General specifications:
- 15-inch color main screen, touch, 1024 x 768 resolution.
- Oxygen drive gas system.
- Automatically compensate for leaks and pipe impedance.
- Support 9 ventilation modes: Manual, VCV, PCV, SIMV-PC, SIMV-VC, SIMV-VG, CPAP/PS, ACGO, PCV-VG.
- Monitor 10 parameters, 3 waveforms, 3 loops.
- Alarm setting: High/Low for parameters: Vt, Paw, MV, FiO2, Apnea.
- Emergency Oxygen Flush Button.
- Mechanical O2-N2O Link System (O2-N2O Link system)
- N2O auto-off function when O2 pressure is lower than 100kPa.
- The mechanical ACGO switch supports ACGO mode.
- 15L/min auxiliary flow column (Air/Oxygen).
- Electronic master flowmeter. Accuracy ±10 % of true value.
- APL quick relief valve
- Support 2 evaporator connection ports, Selectatec / Plug-in standard.
- Support anesthetic agents: Halothane, Isoflurane, Sevoflurane, Desflurane.
- Dimensions (mm): (HxWxD) 1370 x 780 x 690.
- Weight: 145kg.
- Dimensions of the workspace (mm): 545 x 310 x 850.
- Top shelf size: 1635 cm2
- Wheel diameter: 125mm, with central locking
- There are 3 drawers with lock, size 130mm high, 415mm wide, 320mm deep.
- LED worklight: Off, Low, High.
- Power supply:
220-240 Vac, 50/60 Hz, 6A max.
100-240 Vac, 50/60 Hz, 7A max
- Backup battery: Li-Ion, 10.95 VDC, 4.5Ah. Use at least 90 minutes.
- Output of auxiliary sockets (up to 4): 3A.
- Supported ports: RS-232C, DB9, Ethernet (RJ-45), USB, VGA port
Anesthesia machine parameters:
- Support 2 vaporizer ports, Selectatec / Plug-in standard.
- Anesthesia: Halothane, Isoflurane, Sevoflurane, Desflurane.
- Volume of breathing circuit: 2600ml (ventilator) or 1800ml (manual).
- CO2 absorption canister capacity: 1500ml with CO2 bypass function
- Supports 2 types of canister: Pre-pak or Loose-Fill.
- Water trap in the exhalation branch with a capacity of 6ml
- Inspiratory pressure measurement limit: -20 ~ 100 cmH2O. Accuracy: ± 2% FS.
- 2 Flow sensors of differential pressure at the inspiratory and expiratory ports.
- Galvanic-cell Oxygen sensor, FiO2 18%-100%. ± 2.5%. Response ≤ 20s.
- Diameter of connectors: 22mmOD/15mmID. Anesthesia outlet port: 30mm OD
- Mechanical switch for 2 modes of mechanical ventilation – bag breathing.
- Valve APL SP, 5~70 cmH2O. Accuracy ±3 cmH2O . Responsive ≥ 30 cmH2O.
- All parts of the exhalation end are autoclavable (except the sensors).
- Compliance of the whole respiratory system: ≤ 4 mL/cmH2O.
- Breathing system impedance: ≤ 6 cmH2O (both Automatic and Manual mode).
- System-wide leakage: ≤ 60mL @ 3 kPa.
Ventilation settings:
- Patient Size: Adult, Child, Neonate.
- Tidal volume: 20~1500 mL (volume mode), and 5-1500 mL (pressure mode)
- Pinsp: 5~80 cmH2O (1cmH2O increment).
- Plimit: 10~100cmH2O (1cmH2O increment).
- ∆Psupp 3~60 cmH2O (1cmH2O increment).
- Respiration Rate: 2 ~ 100 bpm (1 bpm increment).
- I:E: 4:1 ~ 1:8 (0.5 increment).
- Apnea (Tip:Ti): OFF, 5% – 60% (5% increment).
- Tinsp: 0.2 – 5 s (increase by 0.1 s).
- Trigger window: 5% – 90% (5% increment).
- Trigger flow: 0.2 ~ 15 L/min (0.5L/min increment).
- Pressure trigger: -20 ~ -1 cmH2O (1 cmH2O increment)
- Exp%: 5% – 60% (increase by 5%).
- Average rate: 2 – 60 bpm (1 bpm increment).
- Tslope: 0.0 – 2.0 s (0.1 s increment).
- Apnea I:E: 4:1~1:8 (0.5 increment).
- ∆Papnea: 3 – 60 cmH2O (1 cmH2O increment)
- PEEP: OFF, 3~30 cmH2O (1cm H2O increase). Electronic Control.
- Inlet gas pressure limit: 280 kPa to 600 kPa.
- Maximum flow rate: 120L/min + Fresh Gas flow.
Monitoring parameters:
- Minute Volume: 0~100 L/min.
- Tidal volume:0 ~ 3000 mL/min.
- FiO2: 18% – 100%.
- Paw: -20 ~ 120 cmH2O.
- PEEP: 0 ~ 70 cmH2O.
- Resitance (R): 0 ~ 600 cmH2O/(L/s).
- Compliance (C): 0 ~ 300 ml/cmH2O.
- Rate: 0 ~ 120 bpm.
- Trend: 48 hours.
- Event log: 500 events.